In This Store
Category:Laboratory > Analytical Instrument > Chromatography
Product Name:Unique AutoPure - Protein purification system
Price(USD):Negotiable
Company:Inscinstech Co., Ltd
Factory Location: Building13, 998 PangNan Road, Wujiang District, Suzhou City, Jiangsu Province, CN
Main Sales Markets: Western Europe,Eastern Europe,Asia
Delivery Lead Time: 6-8weeks
Sample Provided: no
Payment Terms: T/T
Unique AutoPure-Protein purification system
The protein purification system is mainly used in the biopharmaceutical fields such as monoclonal antibody, recombinant protein, vaccines, biochemical drug, antibiotics, natural products and polysaccharides. Used for separation and purification of biological products in downstream process
Product Features and Advantages
· Modular design providing multiple configurations to meet special process requirements
· High-precision double plunger infusion pump with multiple patented technologies to ensure excellent separation reproducibility
· All parts of the whole system in contact with liquid are made of biocompatible materials, which meet the requirements of FDA, USP VI and other relevant laws and regulations
· DAD full spectrum direct reading detector to avoid the instability and high failure rate brought by the traditional mechanical grating detector
· The fixed wavelength detector lamp is a long-life LED lamp with a service life of more than 8000 hours, which ·reduces the maintenance cost
· Patented ultraviolet detector flow cell, with low dead volume and low peak expansion, which improves chromatography resolution
· High sensitivity on-line pH, conductivity and UV detector, low dead volume and low peak expansion, which improves chromatography resolution
· Centralized automatic components such as buffer selection valve, automatic sampling valve and column position valve, which improves the automation degree of purification
· Pre-column, post-column, differential pressure alarm and bubble sensor detection functions, which greatly protect the safety of the system and chromatography column
Unique CDsystem chromatography workstation software
Unique CDsystem is a powerful, advanced and stable chromatographic data management system
· CDsystem is integrates the functions of instrument control, method edit, evaluation, standard curve, report editing, user authority management, audit trial, electronic signature, network database etc
· CDsystem meets the requirements of FDA 21 CFR, Part 11 / GLP / GMP, and meets the requirements of the pharmaceutical industry for data traceability and safety
Product Model and Technical Parameters
· Standard configuration
Unique AutoPure Piston Pump
Unique AutoPure25
Piston Pump: Binary gradient biological inert infusion pump
Flowrate Range: 0.001-25ml/min
Gradient type: linear, isocratic, step gradient
Unique AutoPure100
Piston Pump:Binary gradient biological inert infusion pump
Flowrate Range: 0.001-100ml/min
Gradient type: linear, isocratic, step gradient
Unique AutoPure150
Piston Pump: Binary gradient biological inert infusion pump
Flowrate Range: 0.001-150ml/min
Gradient type: linear, isocratic, step gradient
Unique AutoPure300
Piston Pump:Binary gradient biological inert infusion pump
Flowrate Range: 0.01-300ml/min
Gradient type: linear, isocratic, step gradient
Automatic sampling valve
Controlled by software, supports Load/Inject/Waste
Fixed Wavelength Detector
LED, fixed single wavelength, optional 280nm, 260nm, or 254nm
Temperature Detector
Reading range: 0-100 ℃, temperature compensation for conductivity and pH
Conductivity Detector
Detection range: 0-999.999mS/cm, including temperature compensation
Back Pressure Valve Module
20psi (2ml/min)
Inline Filter Module
20μm titanium alloy filter
Control System
Unique CDSystem chromatography software (Without Computer)
· Optional Configuration
pH Detector
Detection range: pH 0-14, temperature compensation
Bubble sensor
Monitor sample loading and air bubbles in the system
UV detector
Fixed single wavelength, fixed dual wavelength, optional 2-4 wavelengths
Inlet Options
Both A/B pumps can be equipped with buffer inlets for 2, 6, 8, or 10 channels
Outlet Options
Optional with 2, 8, and 10 channel large volume outlets
Fraction Collectors
Multiple options such as open, closed, and temperature control fraction collectors
Column Valve
Single column valve, three column valve, including upflow and downflow
Pressure Sensor
System pressure, pre-column and post-column pressure
Sample Pump
Flowrate of 25, 100, 150, or 300ml/min, with optional inlets for 2, 6(only for 300 ml/min), 8, or 10 channels
Contact Us
Tel: (+86) 400 610 1188
WhatsApp/Telegram/Wechat: +86 13621645194
Follow Us: