Basic Info
  • Factory Location: Kyoto,Japan

    Main Sales Markets: Western Europe,Asia,Middle East

  • Sample Provided: yes

    • Hassle-free and safe measurement
    • Easy check of measurement cell
    • No air bubbles, no contamination
    • Calibration at one temperature
    • Easy data transfer
    • Connection to barometer (Option / NON-CE)

    Applications

    Beverages, foods, petroleum, pharmaceutical, alcoholic beverages, etc.

    Specifications

      DA-650 DA-645 DA-640
    Measurement Method Resonant frequency oscillation
    Measurement Range 0 - 3 g/cm3
    Measurement Temp Range 0 - 96℃ (32 - 204.8°F)
    Accuracy*1 Density ±2 x 10-5/gcm3
    (±0.00002 g/cm3)
    ±5 x 10-5g/cm3
    (±0.00005 g/cm3)
    ±1 x 10-4g/cm3
    (±0.0001 g/cm3)
    (Calibration with air and water required.)
    Temp ±0.02°C
    (±0.04° F)
    ±0.03°C
    (±0.05° F)
    ±0.05°C
    (±0.09° F)
    Repeatability *2 Density SD 5 x 10-6g/cm3 SD1 x 10-5g/cm3 SD 5 x 10-5g/cm3
    Minimum Sample Required 1) Approx. 1mL (Syringe) 2) Approx. 2mL (Pump)
    Measurement Time 1) 1 to 4 mins (Manual) 2) 2 to 10 mins (Auto)
    Display

    1) 5.7-inch colour TFT LCD; 640 x 480
    2) Shows density, specific gravity, oscillation frequency, 
    temperature, concentration and other messages.

    Viscosity Correction Yes
    Sampling 1) Manual by syringe
    2) Auto by peristaltic pump
    Method Saves up to 100 different methods in built-in memory
    Stability Four modes of stability according to measurement accuracy and time
    Density Auto Correction

    1) Saves conversion table or formula at your desired temperatures 
    according to your samples.
    2) Temperature conversion table programmed according to ASTM
    standard for petroleum, petroleum products and lubricating oils.

    Auto Conversion 1) Between concentration and density
    2) Between temperature and density
    Statistics 1) Auto or manual calculation of mean value, SD and coefficient
    2) Recalculation, data deletion
    Interfaces 1) LAN:
    x 1; Personal computer (PC)
    2) USB 1.1:
    x 2; USB flash drive, keyboard, 
    barcode reader, Epson inkjet printer *3
    3) RS232C:
    x 2; Dot Matrix Printer, 
    Auto Clean and Sampling Unit,
    Multiple Sample Changer
    Options

    1) Printer : DP-600, IDP-100
    2) Sampling unit, Changer : DCU-551N/H, CHD-502N/H/C
    3) Software : SOFT-CAP (Data Acquisition Software)

    Data I/O 1) USB flash drive as data storage medium
    2) Application Notes provided in USB flash drive
    Wetted Materials PTFE, borosilicate glass, SUS304
    Ambient Conditions 1) Temperature : 5 – 35℃ (41℃ - 95°F)
    2) Humudity : 85%RH or below (No condensation allowed)
    Power Supply AC100 – 240V; 50/ 60Hz (Comes with AC adapter)
    Power Consumption 40W (max. 120W, min.20W)
    Dimensions 320 (W) x 365 (D) x 250 (H) mm
    (12.6 (W) x 14.4 (D) x 9.8 (H) inches)
    Weight 18kg (39.7Ibs)
    Export Packing in Double Carton Box G/W 21kg; 540 (W) x 480 (D) x 460 (H) mm
    (G/W 46.3Ibs; 21.3 (W) x 18.9 (D) x 18.1 (H) inches)
    (May vary in some cases)

    *1, *2:According to KEM standard measurement conditions
    *3:Enquire for applicable models

Send your message to this supplier
  • From:
  • To:
    KYOTO ELECTRONICS MANUFACTURING (Shanghai) Co., LTD.
  • Send a Copy to this Email
  • Message:
    Upload Images / Files
    - Supports jpg, jpeg, png,
     gif, pdf, doc, docx,
     xls, xlsx, txt, rar and zip
    - Max upload 3 files;
     Max  total size: 3MB
    (0/3)

    Enter between 20 to 4,000 characters.This is not what you are looking for ? Post a Sourcing Request Now

  • Verification:
PharmaSources Customer Service