Basic Info
  • Factory Location: Kyoto,Japan

    Main Sales Markets: Western Europe,Asia,Middle East

  • Sample Provided: yes

    Features

    • Titrant Information is stored in an IC chip in the burette unit (Smart burette)
    • Information stored in the electrode cable (Smart electrode)
    • New burette unit
    • Large color TFT-LCD with a touch panel(8.4-inch)
    • Two different user levels and permisions
    • Two different potentials , such as pH/ transmittance , pH/ conductivity or pH/ temperature can be logged simultaneously.
    • Monitoring temperature during titration ensures safe measurements
    • Measurement results are converted to PDF and can be stored in a USB flash drive

    Applications

    Industrial products , Medical goods , Food , Petroleum products , etc.

    Specifications

    Type Automatic Potentiometric Titrator
    Model AT-710S AT-710M
    Product configuration MCU-710S + AT-710 + IDP-100 + Propeller stirrer MCU-710M + AT-710 + IDP-100 + Propeller stirrer
    Detection range
    1) Potentiometric : -2000mV to +2000mV
    2) pH : -20.000 to 20.000pH
    3) Temp. : 0 to 100°C
    Titration mode

    Auto Titration , Auto Intermit , Intermit , Stat , Petroleum Titration , COD

    Method Standard method 120 , Combined method 10 (Max 5 methods can be linked)
    Kinds of titration Potentiometric (acid/base , redox , precipitation) , Photometric , Polarization , Conductivity
    Titration form Full titration (Auto EP detection) , EP Stop , Level Stop , Intersect , EP Stop/Level Stop
    Special application Measurement of electrode potential (pH , potential) , Acid dissociation constant (pKa)
    Simultaneous recording of 2-way input potential (e.g. Titer vs. pH+%T , Titer vs. pH+μS) , Learn
    Key operation Touch panel
    Displays

    1) 8.4-inch color LCD 800 x 600 dots

    2) English / Japanese / Mandarin Chinese / Korean / Russian / Spanish / German / French
    3) 1-channel display 3) Simultaneous 4-channel display (Can also display
    Karl Fischer Moisture titrator simultaneously)
    Calculation Concentration of content , statistics data processing (mean , SD and RSD) and automatic averaging of blank and factor value
    Data storage 500 samples
    GLP conformance

    Registration of operator / User group administration

    Titrant : Reminder of date of factor measurement / Alarm to indicate remaining reagent / Reminder of piston replacement date / Reminder of reagent replacement date / History of factor measurement
    Check performance : Reminder of scheduled checkdate / Record of check results
    Management of electrode : Reminder of calibration date / Record of calibration history / Electrode check / History of electrode check
    Verification of burette capacity : Verification / Record of verification results
    Management of conduction time : Display of operating time
    Burette size 20mL glass burette with brown cover (Standard)
    Optional burette units : 10mL , 5mL , or 1mL
    Burette accuracy 50mL burette(Auto dispenser) +/- 0.5mL
    20mL burette +/- 0.02mL ; reproducibility +/- 0.01mL
    10mL burette +/- 0.015mL ; reproducibility +/- 0.005mL
    5mL burette +/- 0.01mL ; reproducibility +/- 0.003mL
    1mL burette +/- 0.005mL ; reproducibility +/- 0.001mL
    Preamplifier
    1) STD : pH (mV) and mV , 2 inputs (Standard)
    2) PTA : pH (mV) , mV and photometric , 3 inputs
    3) POT :

    pH (mV) , mV and polar , 3 inputs

    4) CMT : pH (mV) , mV and conductivity , 
    3 inputs (factory setting required)
    5) TET : pH (mV) 2 ways and mV , inputs (factory setting required)
    External I/O
    RS-232C port x3 : for Dot matrix printer/Electronic balance/Data Capture Software (SOFT-CAP)
    SS-BUS x1 : for Multiple sample changer , APB
    ELE. x1 : for Smart electrode
    USB x1 : for USB flash drive , Thermal printer , A4 printer , Keyboard , Barcode reader , Foot switch , USB HUB
    TEMP.COMP. x1 : Input terminal for temperature sensor to correct reagent volume , sensor Pt100 , temperature reading accuracy : +/- 0.5°C (burette 1 only)
    -
    LAN ×1 : for Personal computer (PC)
    Extensibility - Measuring instrument : 
    Automatic Potentiometric Titrator (AT-710) , Karl Fischer Moisture Titrator (MKV-710/MKC-710) ; Three of these instruments can be added.
    Automatic piston burette : 
    Can control max 10 burette drives (Including two built-in burette drives)
    Multiple sample changer : CHA-600 , CHA-700
    Ambient condition
    1) Temperature : 5 to 35°C
    2) Humidity : 85%RH or below (no condensation)
    Power source AC100 to 240V +/- 10% 50/60 Hz
    Power consumption
    Main unit : Approx. 30W
    Printer : Approx. 7W
    Dimensions
    Touch panel controller : 225(W) x 190(D) x 42(H)mm
    Titration unit : 141(W) x 296(D) x 367(H)mm (not incl. tubing)
    Printer : 106(W) x 180(D) x 88(H) mm
    Weight
    Touch panel controller : Approx. 1.5kg
    Titration unit : Approx. 4.0kg
    Printer : Approx. 0.4kg
    Conformity standard CE marking EMC : EN61326-1 LVD : EN61010-1 RE Directive
    Burette unit EBU FCC Part15 SubpartC FCC ID : 2ABSVEBU01
Send your message to this supplier
  • From:
  • To:
    KYOTO ELECTRONICS MANUFACTURING (Shanghai) Co., LTD.
  • Send a Copy to this Email
  • Message:
    Upload Images / Files
    - Supports jpg, jpeg, png,
     gif, pdf, doc, docx,
     xls, xlsx, txt, rar and zip
    - Max upload 3 files;
     Max  total size: 3MB
    (0/3)

    Enter between 20 to 4,000 characters.This is not what you are looking for ? Post a Sourcing Request Now

  • Verification:
PharmaSources Customer Service